Simulator fidelity

The ResistaMet test suite uses a stateful in-memory Keithley 2400-series simulator (tests/fakes/fake_keithley.py) so the full SCPI logic surface can be exercised in CI without a real instrument. This document records:

  1. Which simulator behaviors are validated against captured hardware traces.
  2. Which behaviors are intentionally simplified (and what tests cover the gaps via real hardware).
  3. How to recapture the golden traces when firmware drifts or new scenarios are needed.

Reference instruments

Captures, quirk validation, and trace replay were performed against two physical instruments from the Keithley 2400 family:

Field Primary (capture source) Cross-model validation
Model Keithley 2420 (3 A SourceMeter) Keithley 2400 (1 A SourceMeter)
Serial 1230523 1175680
Firmware C30, March 17 2006 C30, March 17 2006 (identical)
Option codes /H/L /K/J
Interface GPIB-USB-HS, address 24 GPIB, address 3
Line frequency 60 Hz 60 Hz
DUTs 100 Ω (99.53 Ω), 10 kΩ (9914 Ω), 1 MΩ (1.029 MΩ) 100 Ω (99.50 Ω)

Cross-model evidence: every committed SCPI fixture (29 traces across 100 Ω / 10 kΩ / 1 MΩ DUTs) and every quirk-trigger test captured from the 2420 reproduces byte-equivalent (configuration queries) and within 5 % / exact-on-compliance-bit (measurement queries) on the 2400 — 29/29 fixtures and 6/6 quirks pass cross-model (validated 2026-04-30). Cross-model pass counts by operating regime: 15/15 at 100 Ω (mA / 0.1 V), 8/8 at 10 kΩ (100 µA / 1 V, including the 2-wire and negative-source-V variants), 6/6 at 1 MΩ (µA / 1 V). The two instruments share firmware C30 but differ in model, serial number, current rating, and option codes; the fixtures and the simulator are therefore validated against the 2400 family across the full operating envelope of the production code, not only one specific unit at one operating point.

Three decades of resistance are exercised:

  • 100 Ω at 1 mA / ~0.1 V — lowest current range, source-V mode primarily
  • 10 kΩ at 100 µA / ~1 V — middle range, plus 2-wire and negative-source-V variants
  • 1 MΩ at 1 µA / ~1 V — lowest source-current range, exercises the µA sense regime

The simulator passes every trace without per-DUT or per-range code changes — the trace-replay test only pins the STAT compliance bit, so the per-mode STAT baselines that differ across ranges don't trigger mismatches.

The simulator targets the broader 2400/2410/2420/2425/2430/2440/2450 family. Bits encoded in the FORM:ELEM STAT element (notably the compliance bit at position 3) are 2400-series specific; the 2450 may differ.

What the simulator models

Validated against captured traces (tests/fixtures/scpi_traces/) and against live hardware quirk triggers:

  • Default state after *RST: 33 settable properties match the real instrument's documented and observed defaults (line frequency, sense function, compliance limits, source ranges, output mode, etc.). Verified by baseline_reset_state.json.
  • Five measurement modes (resistance, source‑V, source‑I, four‑point probe, voltage sweep): full configuration round‑trips and :READ? element layouts match. Verified by per‑mode trace fixtures.
  • Source value echoing under compliance: in source-V mode, the VOLT element of :READ? echoes the programmed setpoint even when the instrument is internally clamping current at the compliance limit. The CURR element reflects the actual clamped current. Symmetric for source-I mode. Verified by compliance_v_in_compliance.json.
  • STAT compliance bit: bit 3 of the STAT element of :READ? is set iff the source is in real compliance. (Note: this is not the same bit as the +114 Measurement Event Register entry at bit 14 documented in the manual's Table B-1 — the FORM:ELEM STAT element uses a different bit layout. The simulator has _STAT_BIT_COMPLIANCE = 1 << 3 and per-mode baselines empirically observed on the 2420.) Verified by tests/hardware/test_quirk_triggers.py::test_compliance_bit_is_bit_3.
  • FORM:ELEM canonical re-ordering quirk: argument order in :FORM:ELEM is silently re-ordered by the instrument; argument set is honored. Verified by quirk_form_elem_canonical_order.json.
  • Auto-ohms rejection (error 825): after :SENS:FUNC 'RES', attempts to write :SOUR:CURR, :SOUR:CURR:RANG, or :SENS:VOLT:PROT queue error 825 "Invalid with auto-ohms on" until :SENS:RES:MODE MAN is sent. Verified by quirk_auto_ohms_rejects_source.json.
  • :INIT:CONT undefined header (error -113): the :INIT:CONT subsystem does not exist on the 2400 series; :READ? performs init + trigger + fetch atomically. Verified by quirk_init_cont_unsupported.json.
  • Concurrent measurements default ON: after *RST, :SENS:FUNC:CONC is 1, costing 3× measurement time unless explicitly disabled. Verified by quirk_concurrent_default_on.json.
  • Hardware averaging filter: :SENS:AVER, :SENS:AVER:TCON, :SENS:AVER:COUN round-trip values correctly. Verified by filter_repeat_x10.json.
  • Offset-compensated ohms: :SENS:RES:OCOM ON/OFF round-trip and configuration sequence. Verified by offset_compensated_ohms.json.
  • Output-off HIMP mode: :OUTP:SMOD HIMP/NORM round-trip. Verified by output_off_himp.json.
  • Sweep engine: :SOUR:VOLT:MODE SWE plus :SOUR:VOLT:START/STOP/STEP plus :TRIG:COUN N causes one :READ? to return N consecutive (V, I, STAT) triples. Verified by sweep_v_0_to_0p5v_5pts_into_100ohm.json.
  • Error queue FIFO: :SYST:ERR? returns the oldest queued error in <code>,"<message>" format with no leading + on positive codes; 0,"No error" when empty.

What the simulator does NOT model

These gaps are documented here so reviewers know where the hardware tier provides the only safety net.

  • Real measurement noise: the simulator computes V/I from dut_resistance_ohms via Ohm's law. Real readings exhibit noise on the order of 0.05% at NPLC=1 on the 2420. Tests that compare simulator output to captured hardware traces use a 5% tolerance for V/I and only require exact agreement on the STAT compliance bit.
  • NPLC and source-delay timing: the simulator returns :READ? responses immediately. It does not delay by (NPLC × line_period) or honor :SOUR:DEL. Tests that need timing fidelity must run on real hardware.
  • Filter time integration: :SENS:AVER ON is recorded as a state flag but does not actually average multiple internal samples; the simulator returns the same single computed value either way.
  • Range-dependent precision and clamping: real instruments quantize to range-dependent step sizes and reject inputs outside the active range. The simulator accepts arbitrary floats.
  • Thermoelectric EMF: configurable via dut_voltage_offset, but the simulator does not introduce drift or temperature dependence.
  • :STAT:MEAS:COND? and other status-register subsystems: the simulator returns 0 (a placeholder); tests that need the Measurement Event Register layout must run on real hardware.
  • Front-panel state, calibration data, save/recall slots: ignored by the simulator.
  • Delta mode hardware capture: the production worker's current- reversal path (+I → :READ? → -I → :READ? → +I) is exercised in CI by test_workers.py::test_four_point_delta_mode_alternates_polarity, which asserts the right SCPI command sequence is sent. We do not currently ship a hardware trace for delta mode; an attempt at 10kΩ exposed a polarity asymmetry in the lab Kelvin wiring that was wiring-specific (current direction reversal hit voltage-protection compliance) rather than instrument-specific, so the trace was not committed. A future capture on a known-symmetric setup would close this gap.

Test tier overview

Tier Hardware required Files Pass count
Pure unit No test_buffers.py, test_calculations.py, test_config.py, test_data_export.py, test_system_utils.py, test_widgets.py (existing)
GUI smoke No (offscreen Qt) test_gui_smoke.py (existing)
Trace replay No test_fake_matches_hardware.py 30/30 (29 traces + 1 sanity)
SCPI wrapper No test_instrument.py 21/21
Worker integration No test_workers.py 17/17
Hardware quirks Yes tests/hardware/test_quirk_triggers.py 6/6 on 2420 + 6/6 on 2400 (validated 2026-04-30)
Hardware recapture Yes tests/hardware/test_recapture_traces.py 29/29 on 2420 + 29/29 cross-model on 2400 (validated 2026-04-30); set RESISTAMET_DUT_OHMS to filter

The CI pipeline (.github/workflows/) runs the four no-hardware tiers on every push. The two hardware tiers run locally before each release with the bench instrument connected:

RESISTAMET_HARDWARE_ADDR=GPIB0::24::INSTR pytest tests/hardware/ -v

Community cross-model submissions

Anyone with a Keithley 2400-family instrument can run scripts/community_capture.py against a 100 Ω, 10 kΩ, or 1 MΩ 4-wire Kelvin reference DUT and submit the resulting SCPI traces via the Keithley compatibility issue template. Accepted submissions land under tests/fixtures/scpi_traces_community/<model>_<serial>/ and are automatically replayed through the simulator by tests/test_community_traces.py on every CI run.

When a submitted trace fails the simulator, that's evidence of real cross-model variance — the maintainers either update the simulator to handle it (with the trace as the regression fixture) or document the divergence here as a known gap. Either way, the testable surface grows.

The model spec table in resistamet_gui/instrument.py (mirrored in scripts/community_capture.py so the script can run without the project installed) is updated as new models are confirmed by submissions.

Recapturing golden traces

When firmware changes or new measurement scenarios need coverage:

  1. Wire the reference DUT (default: 100 Ω resistor in 4-wire Kelvin).
  2. Run the capture script with the instrument connected:
python tests/hardware/capture_traces.py

New JSON fixtures appear under tests/fixtures/scpi_traces/.

  1. Re-run the simulator validation:
pytest tests/test_fake_matches_hardware.py -v
  1. If the simulator now diverges, update tests/fakes/fake_keithley.py to match the new captures, then iterate until both test_fake_matches_hardware.py (CI) and tests/hardware/ (bench) pass.

Why this design

pyvisa-sim, the standard simulation package, supports static SCPI dialogues but cannot easily produce computed responses (:READ? values based on the configured source mode and DUT model, sweep response synthesis, the auto-ohms quirk's conditional error queueing). A custom stateful fake bypasses the YAML format entirely while preserving the PyVISA ResourceManager.open_resource interface, so the production code under test runs unchanged in CI.

Captured hardware traces rather than hand-written expected values means a simulator regression shows up as a divergence from a recorded instrument dialogue, not from an opinion about correct behavior. The trade-off: updating the simulator requires bench access to re-capture, not just an edit to a fixture file.