Settings reference

Open via the Settings button or menu. Four tabs: Measurement, Display, File, Output. All values are persisted per-user-profile in config.json (gitignored, in the working directory). Cross-references in this page link to Concepts for term explanations.

Measurement tab

General instrument

Setting Default Range / type What it does
GPIB Address GPIB0::24::INSTR VISA resource string Where to find the instrument. Detect Devices scans all available VISA buses and offers the result in a picker. Machine-local: stored keyed by hostname rather than in the shared profile, so a NAS-shared config.json works across lab PCs with different wiring. Any legacy single-address slot is auto-lifted on first save.
Sampling Rate 10.0 Hz 0.1 – 100 Hz Default readings per second. Each measurement tab has its own spinbox that overrides this. Each per-tab spinbox is dynamically capped based on the active (NPLC × auto-zero × filter × offset-comp) combination; if you commit a value above the cap the status bar reports the cap and suggests the cheapest single setting change to reach the requested rate. See Concepts → Rate-cap predictor.
NPLC 1 0.01 – 10 Integration time per reading. Lower = faster + noisier; higher = slower + cleaner.
Settling Time 0.2 s 0 – 10 s Delay after output enable before the first reading. Lets DUT + instrument stabilize.

Resistance defaults

Applied to the Resistance tab on launch; the tab itself has live widgets that override these per-run.

Setting Default Notes
Test Current 1 mA DC current sourced through the DUT
Voltage Compliance 5 V Maximum allowed across the DUT
Measurement Type 2-wire Or 4-wire for Kelvin

Voltage Source defaults

Setting Default Notes
Source Voltage 1.0 V Accepts negative values
Current Compliance 100 mA
Duration (hours) 0 (= run until stopped) Set non-zero for unattended runs

Current Source defaults

Setting Default Notes
Source Current 1 mA Accepts negative values
Voltage Compliance 5 V
Duration (hours) 0 (= run until stopped)
Stop on compliance False Auto-Stop when compliance is reached. Useful for protecting sensitive samples in source-V too.

Advanced instrument settings

Setting Default Notes
Enable Hardware Filter True Use the Keithley's built-in averaging filter (see Concepts → Hardware averaging filter)
Filter Type repeat repeat (N readings → 1 result, then repeat) or moving (running average)
Filter Count 5 1 – 100
Enhanced accuracy in Resistance mode True Offset-compensated ohms. See Concepts → Enhanced R mode.
Touch-safety warn threshold 30 V 0 – 1100 V. Voltage at or above this triggers the warning modal before run start. 0 disables. See Concepts → Touch-safety warning.
Suppress touch-safety warning for this profile False Equivalent to clicking "Don't show again" on the modal. Uncheck to re-enable.

Auto-Zero lives on each tab, not in Settings

Pre-v1.10 it was here, but per-tab makes more sense: 4PP and vdP force auto_zero='on' regardless; sensor modes (resistance, source_v, source_i) expose the knob on the tab itself so re-tuning the speed/accuracy trade doesn't require a Settings dialog detour.

Display tab

Setting Default Notes
Enable Real-time Plots True Disable for high-rate runs (>50 Hz) where rendering eats CPU
Resistance Plot Color #D55E00 (Wong vermillion) Matches the V/I/R/P live-readout label colors
V Source Plot Color #0072B2 (Wong blue)
I Source Plot Color #009E73 (Wong green)
Data Buffer Size (points) 0 (Unlimited) 0 = no cap. pyqtgraph downsamples on render (peak-mode + clipToView) so a 17-hour / 270k-sample run is ~13 MB and stays smooth. Set a finite cap only if RAM is tight.

Pre-v1.10 buffer-size default was 1000

Old configs that explicitly set buffer_size to a small number silently truncated the live trace on overnight runs. New default is unlimited; existing configs override unless you reset.

File tab

Setting Default Notes
Auto-save Interval 60 s How often to flush data to disk during a measurement. Lower = less data loss on crash, but more I/O.
Data Directory measurement_data Root for run output. A subdirectory is created for each user. Browse... button opens a folder picker.

Output tab

See Data Outputs for the full format reference; this tab just selects which backend to use.

Setting Default Notes
Output format csv One of: csv (default, single .csv with #-prefixed metadata header), hdf5 (requires h5py), csv+legacy_json (pre-2.0 dual emit for back-compat with downstream pipelines)
Compression never Applies only to the csv backend. never / always / auto. Default is never because Excel, Origin, and most lab tools can't open .csv.gz directly. HDF5 is always internally gzip-compressed regardless.
Auto-compression threshold 5 MB Only used by the auto compression policy

HDF5 grayed out?

If h5py isn't installed, the HDF5 option is disabled with a tooltip explaining why. pip install h5py and restart to enable it.

Configuration storage

Settings are kept in config.json in the working directory (gitignored — per-user, not portable). Structure mirrors the four Settings tabs:

{
  "measurement": {...},
  "display": {...},
  "file": {...},
  "output": {...},
  "users": ["alice", "bob"],
  "last_user": "alice",
  "user_settings": {
    "alice": {"measurement": {...}, ...},
    "bob": {"measurement": {...}, ...}
  }
}

When a user is selected, their per-user overrides deep-merge on top of the global defaults. Editing this file by hand works but is error-prone — prefer the dialog. If the file gets corrupted, delete it and ResistaMet will recreate from DEFAULT_SETTINGS (in resistamet_gui/constants.py).

Profiles menu

The top-level Profiles menu has two entries that operate on the currently-active tab (Resistance / Voltage Source / Current Source / Four-Point Probe — Sweep and vdP are excluded):

  • Save Profile for Current Mode... writes the active tab's measurement settings to a .json file (suggested name <mode>_profile.json). Useful for "sensor A typical run" vs "sensor B typical run" templates that you swap between samples.
  • Load Profile to Current Mode... reads a .json and applies known fields to the active tab's widgets. Unknown fields are ignored. Legacy fpp_thickness_cm is auto-converted to fpp_thickness_um for backward compatibility.

These are file-level profiles distinct from the per-user settings stored in config.json. They don't move the GPIB address (which is machine-local).

Results Viewer tab

A 7th tab "Results Viewer" is included by default. It lets you:

  • Open any v2.0 .csv or .csv.gz file produced by ResistaMet GUI (also accepts older legacy CSVs)
  • Pick any column from the dropdown to plot vs elapsed_s
  • Read the trailing metadata (run mode, sample name, start time) into the status log

No write-back — it's strictly a post-run viewer.

Per-tab settings (not in this dialog)

Some settings live on the individual mode tabs because they're per-measurement-context, not session-wide:

  • Auto Zero — Resistance / Voltage Source / Current Source tabs (each)
  • Enhanced accuracy — Resistance tab (separate from the global Settings checkbox, which sets the default)
  • Source Range Auto / Voltage Range Auto — every tab
  • Cable null — Resistance tab
  • Delta mode (current reversal) — Four-Point Probe tab Advanced section
  • Geometry / dopant / temperature — Four-Point Probe tab (drive F84 corrections)
  • Sample thickness — Four-Point Probe and Van der Pauw tabs
  • Sweep start / stop / step / direction / per-step delay — I-V Sweep tab

These follow the same per-user persistence — they're written into the user's section of config.json when the user clicks Save Settings on the active tab.